Thermal Measurements

Thermal measurements are used in a variety of applications involving LEDs, laser diodes and VCSELs. The measurements techniques fall in to two broad categories: measurements of the device’s junction temperature, and determination of thermal resistance including measurements of the thermal path from the junction to an exterior point. These thermal measurements are used by solid-state device scientists to understand and improve devices, by test engineers when making optical measurements, and by luminaire / fixture designers to understand the thermal performance of lighting products.

Easily measure junction temperature with Vektrex SMU

What thermal measurements are possible when using Vektrex SMU?

The LED and laser thermal measurements possible using Vektrex pulsed SMU with the optional BIAS function include:

In-situ Tj Measurement for Individual LEDs or Luminaires

The Junction Temperature (Tj) is the operating temperature of the LED or laser diode’s semiconductor junction. When operating, the Tj of a semiconductor is higher than the surrounding ambient environment, and the heat dissipated in the semiconductor flows through a path that includes the various packaging components such as solder, MCPCB, and heat sink. This is illustrated in the diagram below.

Typical LED Heat Flow Path

Tj has a direct impact on the long-term reliability of a laser or LED. Higher in-situ Tj leads to reduced long-term reliability. At temperatures above 150 ˚C the device may fail immediately. Tj also impacts the operating characteristics of LEDs and lasers. For example, an amber LED’s light output decreases dramatically as junction temperature increases.

LED manufacturers typically provide performance specifications at a specific Tj along with derating factors that allow performance at other temperatures to be calculated. Using these values, along with actual in-situ measurements, product designers verify the product will perform within its advertised specifications.

The in-situ junction temperature is the Tj of the device (e.g., an LED) in its operating environment. For example, the operating environment might be a burn-in fixture, an enclosed luminaire, or any other environment where the device is installed. 

In-situ Tj can be measured in two ways: directly, with the Electrical Test Method (ETM); or indirectly using the estimated thermal resistance of the heat path. Estimating Tj via thermal resistance can be risky though – if the actual Tj is incorrect, the product could fail, perform poorly, or it might be overdesigned and more costly than it needs to be.

The ETM method relies on a high-duty-cycle heating current and fast transition to measurement current. A typical In-Situ LED junction temperature current waveform when using the ETM shows this fast transition.

TJ Waveform
Vektrex Sample Junction Temperature Waveform

There are many reasons to undertake in-situ Tj measurements. For example:

  • Mass-marketed lighting products must include heat sinks for cooling. With more accurate Tjmeasurements, thermal resistance uncertainty budgets can be reduced, allowing heat sinks to made smaller. This reduces product cost and shipping cost.
  • Many product failures can be traced to excessive junction temperatures or to large swings in the junction temperature. Warranty losses related to a Tj design error can exceed several hundred thousand dollars for a high-volume product.
  • For products used in designs that are produced over many years, knowledge of the accurate Tj is helpful when comparing replacement devices to the original LED or laser.
  • For manufacturers of LEDs and lasers, it is important to provide the most accurate measurement data possible. This data should include the precise Tj at the time of measurement.
  • In-situ Tj measurement is a useful tool for qualifying and adjusting a design for ideal performance as it will expose the effect of small changes in the product. For example, the impact of various thermal interface materials on the heat flow out of a COB package can be measured quantitatively in just a few minutes.
  • In-situ Tj should be compared with the manufacturer’s maximum allowed junction temperature to make sure it is below the maximum allowable Tj. Usually, designers want to see Tj 20-40% below this temperature. This is particularly important for pulsed overdrive operation.

What products are used to measure junction temperature?

Contact Vektrex to get information about the products we offer to get Junction Temperature Measurements.

Tj Change During Photometric Measurements

An LED or laser heats during photometric measurements and thus the junction temperature increases. The increased Tj alters the measurement, reducing accuracy. In high-power situations, the measurement may be useless. Photometric measurement standards, such as the Illuminating Engineering Society’s LM-85, require Tj shift to be assessed and accounted for, either by altering the testing temperature, or by applying a correction to the measurement. LM-85 provides two different methods in its annexes for assessing Tj during measurements. One uses only the primary current source and the other uses both the primary and a secondary bias source.

What products are used to measure junction temperature change during light measurement?

Contact Vektrex to learn more about the tools Vektrex offers to assess junction temperature change during light measurement.

Pulsed Operation Tj Rise Assessment

An LED or laser being operated with pulsed drive will experience both average and transitory heating. When designing luminaires, it is useful to know the average Tj during pulsed operation, and also the peak Tj. Average Tj can be compared with data sheet Tj values when looking at characteristics such as optical power output and wavelength. Peak Tj should be measured to ensure the device is well below the maximum allowed Tj. As discussed in the previous section, the IES LM-85 document includes procedures in its annexes for assessing Tj.

This graph shows LED Tj for 1ms, 120Hz pulsed operation of a red LED. The pre-pulse, post-pulse and average temperature are shown, along with a polynomial model derived from the average Tj plot.

What products measure juntion temperature during pulsed operation?

Contact Vektrex to learn more about how to assess junction temperature rise during pulsed operation.

Thermal Resistance Measurement

Thermal resistance is a parameter that quantifies the resistance heat sees when flowing through a path. It is always expressed as the resistance from one point to another (e.g., from junction to ambient air). The typical units are ˚C/W. LED and laser manufacturers typically provide the thermal resistance for the device itself – that is the resistance from junction to case. However, this does not include the resistance of other structures such as printed circuit boards and heat sinks. In many design situations, such as luminaire design, knowing the thermal resistance of these structures is very useful.

The JEDEC JESD51-14 Transient Dual Interface Method is the best way to measure thermal resistance. This method requires a two-level current source and a voltage digitizer that can sample for long time periods. The LED or laser’s voltage is sampled after a transition and the resulting cooling curve is analyzed to calculate the thermal resistance. JEDEC provides a free software tool that does this; the SpikeSafe Control Panel application captures the data and outputs it in a file format compatible with the JEDEC tool.

LED Cooling Curve Captured with SpikeSafe SMU and Vektrex Control Panel software application.
LED Cooling Curve Captured With SpikeSafe SMU and Control Panel Application

What tools are used to measure thermal resistance?

Contact Vektrex to get more information about the tools available to measure thermal resistance.

Heat Path Analysis

Heat path analysis is similar to thermal resistance measurement, but instead of a bulk parameter representing the entire heat path, the focus of the measurement is usually a specific structure, such as the die attachment or the thermal interface material. To allow the different materials in the heat path to be identified, the cooling curve is transformed into a network model called a structure function. Structure functions for two different conditions can then be compared and the thermal characteristics of the material of interest can be inferred. The special-purpose tool that popularized structure function analysis is called the T3Ster, produced by MICRED.

The calculations that the T3Ster uses to transform a cooling curve into a structure function were placed in the public domain by MICRED and JEDEC in 2010 in the form of a free analysis tool, TDIM Master. However, most labs found it was difficult to obtain cooling curves with sufficient resolution using the general-purpose instrumentation available at the time, because voltage readings were noisy. In addition, the need to change sampling intervals during the cooling time required custom software to implement logarithmic sampling. Often repetitive captures were required, meaning a single measurement could take more than an hour. Combining the captured data was also difficult.

With the addition of logarithmic sampling to its true-differential digitizer, an accurate cooling curve can be collected in seconds with the SpikeSafe SMU with no custom programming. The SMU’s digitizer provides sub-millivolt detail, and it collects the curve in one capture, automatically altering its boxcar averaging on-the-fly. The log sampling feature is available in every SMU model from 500mA to 60A. The SpikeSafe Control Panel application uploads the captured log data and outputs it in a file format compatible with the JEDEC tool. Using the JEDEC tool, structure functions can then be obtained with just a few clicks after importing the Control Panel data.

Example Structure Function Analysis demonstrates damaged thermal pad.
Structure Functions for Luxeon CZ Green LED, Generated With SpikeSafe SMU, Red: Normal, Blue: Damaged Thermal Pad

What tools are needed for heat path analysis?

Contact Vektrex to get information about T3ster-like measurements

What is the difference between JEDEC and sp2 Measurements?

The IES Single Pulse measurement methods, including SP-2, use the LED’s forward voltage to determine Junction Temperature — Tj. The JEDEC 51-series standards also use forward voltage to determine Tj. While both methods use the LED’s voltage temperature sensitivity to infer temperature, there are some differences in the way that is done. These differences arose because LM-85 is focused on optical and electrical measurements — and mostly on the optical properties of the LED. On the other hand, the JEDEC standards are focused on the LED’s thermal properties, such as its operating junction temperature, its thermal resistance and its structure function.

With these priorities in mind, the authors of the two standards made choices that were optimized for their primary document focus — and, in the area of Tj measurement, they chose slightly different methods.

First, let’s start with the JEDEC standards. The JEDEC standards, particularly JESD51-1, introduced the Electrical Test Method (ETM). JESD51-1 is an original reference that does a good job of explaining the method in easy-to-understand terms. The more recent JESD51-51 presents the ETM in greater detail, and it adds the concept of structure functions. JESD51-51 should be part of every lab’s library. Both JESD51-1 and JESD51-14 measure thermal properties during the cooling time – the time just after the LED is turned off. During this time the LED is actually not completely turned off, it is driven by a small measurement current. As the LED cools, its changing forward voltage is sampled, and the resulting data can be used to calculate Tj, thermal resistance, and the structure function.

The JEDEC authors could have chosen to collect the same data during the LED’s heating time, but the current amplitude is much higher then as the LED is being driven a high power. This high current introduces two errors that can be significant in some situations. The first error relates to the applied power. The calculations assume a power step to a constant level, but most test apparatus actually drives the LED under test in constant-current mode. Since the LED’s voltage decreases as it heats, the applied power also decreases. In subsequent calculations that assume constant power, the calculated Tj-increase, especially in later time samples, will be less than the actual. The second error is caused by resistive heating in the conductors in the LED and surrounding circuitry. Copper conductors have a positive temperature coefficient of resistance, and so when they heat, their resistance increases. This increase causes the measured Vf to increase. Again, the effect is greater as time goes on and the error results in calculated Tj-increase values that are lower than actual.

JEDEC’s authors wanted to maximize the accuracy of the thermal measurements and so they chose to make their measurements during the cooling time. This meant that they needed an extra current source — the measurement current (or bias) source, — but that was an acceptable tradeoff to get good data. As mentioned, LM-85’s authors had a different focus — good optical measurements. In this situation, the JEDEC cooling-time approach has several drawbacks: 1) It requires a second current source. 2) It lengthens the measurement time. because the cooling curve data has to be collected after the pulse has ended. 3) Aligning the Tj measurements with the optical measurements is more complex as the two measurements do not occur at the same time.

Choosing heating-time measurements would avoid these drawbacks. In addition, the LM-85 authors realized heating-time errors were not as significant in typical optical measurement applications. The constant-power error was not significant, and since most test setups use 4-wire Kelvin connections, neither was the interconnect wiring error. So going with a heating time measurement was the best choice.

While their use of heating-or-cooling time differs, both the LM-85 and the JEDEC standards calculate Tj(0) the same way, using projections. After the initial Tj(0) though, they differ again. The IES methods are optimized for quick optical measurements — for example the SP methods make use of simple average-voltage measurements that coincide with the optical measurement. The JEDEC methods take logarithmically sampled measurements (sometimes over minutes or hours). This data is then highly processed to determine characteristics such as the LED’s thermal resistance and its structure function.

Products for Junction Temperature Measurements

JD-2020 – The JD-2020 LED Thermal Resistance Analyzer is a high-speed instrument designed to conduct transient/in-line thermal resistance (RΘ) measurements.

MICRED T3STER – The MicReD T3Ster specializes in highly-detailed, fast, accurate thermal testing, measurement, and characterization of IC packages, LEDs, and systems. 

Downloads

Measuring Junction Temperature